ЭС-10-07

 
Ion-selective electrodes
ES-10-07
Ion-selective electrode glass to measure the rNa in demineralized water
ES-10-07
ES-10-07 - an electrode designed to measure the activity concentration of sodium ions in a chemically demineralized water and condensate steam pressure boilers and designed to work with devices rNa-201-205 rNa.
Electrode sealed construction with high stability of parameters. Electrostatic shield protects the electrode from external electric fields.
Cap electrodes are flush with the glass housing.
Parameter ES-10-07
Measuring range rNa at 25 ° C From -0.5 to 7.5 rNa
Temperature of the analyzed medium From 10 to 100 ° C
The coordinates of the isopotential point (for the range pNa: -0.5 to +3 pNa) рNai = 3 рNa, Еi = -22.5 мВ
Electrical resistance at 20 ° C from 150 to 450 MW
Indicator of the electrode Special Glass
Overall dimensions of the electrode, not more than 
- the diameter of the electrode immersion 
- the length of the electrode 
- Cable,

12 мм 
160 мм 
400 мм
Jack Plug cable
The number of electrodes per pack. 2

  • http://zipgomel.by/themes/acquia_marina/images/green-bullet.png); list-style: none; background-position: 0px 2px; background-repeat: no-repeat no-repeat; ">Prices are formed based on the amount, timing and form of payment. The company has a flexible system of price cuts on capital purchases. Regular customers discounts.