И-160.1 МП

Devices Laboratory

LABORATORY IONOMER I-160.1MP

И-160.1МП

Is next device of  series range EV‑74, I‑130, I-160

 

Microprocessing  ionomer I-160.1MP with potentiometric titration function is used to determine  alkali neutralization number in oils according to GOST11362-96 (ISO-6619-88) «Petroleum products and lubricants; neutralization number; potentiometric titration method» is designed to determine in aqueous solutions the activity of hydrogen ions (pH), redox potential (Eh), activity and concentration of ions:

H+, Li+, Na+, K+, NH4+, Ag+, X +, NO3, ClO4, F ,Cl, Br, I, CN, SCN, Ca++, Ba++, Mg++, (Ca+Mg)++, Pb++, Cd++, Cu++, Hg++, X++, CO3– –, S – – etc.

As well as for potentiometric titration when device is completed with additional equipments.

 

Advantages of device

v   Combines two devices.

v   Can be used as ordinary рН-meter-ionomer.

v   Performance reliability, simplicity of operation.

v   Compliance with all domestic and imported ion-selective and рН electrodes.

v   User-friendly functional keyboard.

v   Large and contrast display device.

v   Possible connection to the computer.

Delivery kit

I-160.1MP

I-160.1MP (extended)

Measuring converter

1 pc.

1 pc.

Temperature compensator ТКА-1000.1

1 pc.

1 pc.

Electrode EVL-1М3.1

1 pc.

1 pc.

Microburette

1 pc.

Operational documentation (with electromagnetic valve)

1 kit

1 kit (extended)

Operational documentation

1 kit

1 kit

 

Specifications

 

Measurand

Unit of measurement

Measurement  range

Discreteness

Limits of permissible

main absolute error

Activity of hydrogen ion and other monovalent and divalent cations and anions

рН (рХ)

from –20,00 to +20,00

0,001

±0,02 for monovalent ions

±0,04 for divalent ions

Oxidation-reduction potential

mV

from –3000 to +2000

0,1

±1,0

Temperature of  analyzed environment

ºС

v –20 to +150

0,1

±0,5

AC power supply

V / Hz

(220±22) / (50±0,5)

Overall dimensions

mm

230 х 220 х 85

Weight

kg

2,0